ORIGINAL ARTICLE
Year : 2015  |  Volume : 10  |  Issue : 3  |  Page : 221-228

Accuracy and repeatability of refractive error measurements by photorefractometry


1 Ophthalmic Epidemiology Research Center; Department of Ophthalmology, Shahid Beheshti University of Medical Sciences; Basir Eye Safety Research Center, Basir Eye Clinic, Tehran, Iran
2 Basir Eye Safety Research Center, Basir Eye Clinic; Ophthalmic Research Center, Shahid Beheshti University of Medical Sciences, Tehran, Iran
3 Basir Eye Safety Research Center, Basir Eye Clinic, Tehran, Iran
4 Ophthalmic Research Center, Shahid Beheshti University of Medical Sciences; Department of Epidemiology and Biostatistics, Tehran University of Medical Sciences, Tehran, Iran
5 School of Medicine, Shahid Beheshti University of Medical Sciences, Tehran, Iran

Correspondence Address:
Hamideh Sabbaghi
Ophthalmic Research Center, Shahid Beheshti University of Medical Sciences, No. 23, Paidarfard St., Boostan 9 St., Pasdaran Ave., Tehran 16666
Iran
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/2008-322X.170360

Rights and Permissions

Purpose: To determine the accuracy of photorefraction and autorefraction as compared to cycloautorefraction and to detect the repeatability of photorefraction. Methods: This diagnostic study included the right eyes of 86 children aged 7-12 years. Refractive status was measured using photorefraction (PlusoptiX SO4, GmbH, Nürnberg, Germany) and autorefraction (Topcon RM800, USA) with and without cycloplegia. Photorefraction for each eye was performed three times to assess repeatability. Results: The overall agreement between photorefraction and cycloautorefraction was over 81% for all refractive errors. Photorefractometry had acceptable sensitivity and specificity for myopia and astigmatism. There was no statistically significant difference considering myopia and astigmatism in all comparisons, while the difference was significant for hyperopia using both amblyogenic (P = 0.006) and nonamblyogenic criteria (P = 0.001). A myopic shift of 1.21 diopter (D) and 1.58 D occurred with photorefraction in nonamblyogenic and amblyogenic hyperopia, respectively. Using revised cut-off points of + 1.12 D and + 2.6 D instead of + 2.00 D and + 3.50 D improved the sensitivity of photorefractometry to 84.62% and 69.23%, respectively. The repeatability of photorefraction for measurement of myopia, astigmatism and hyperopia was acceptable (intra-cluster correlation [ICC]: 0.98, 0.94 and 0.77, respectively). Autorefraction results were significantly different from cycloautorefraction in hyperopia (P < 0.0001), but comparable in myopia and astigmatism. Also, noncycloglegic autorefraction results were similar to photorefraction in this study. Conclusion: Although photorefraction was accurate for measurement of myopia and astigmatism, its sensitivity for hyperopia was low which could be improved by considering revised cut-off points. Considering cut-off points, photorefraction can be used as a screening method.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed1359    
    Printed15    
    Emailed3    
    PDF Downloaded179    
    Comments [Add]    
    Cited by others 1    

Recommend this journal