ORIGINAL ARTICLE
Year : 2015  |  Volume : 10  |  Issue : 3  |  Page : 221-228

Accuracy and repeatability of refractive error measurements by photorefractometry


1 Ophthalmic Epidemiology Research Center; Department of Ophthalmology, Shahid Beheshti University of Medical Sciences; Basir Eye Safety Research Center, Basir Eye Clinic, Tehran, Iran
2 Basir Eye Safety Research Center, Basir Eye Clinic; Ophthalmic Research Center, Shahid Beheshti University of Medical Sciences, Tehran, Iran
3 Basir Eye Safety Research Center, Basir Eye Clinic, Tehran, Iran
4 Ophthalmic Research Center, Shahid Beheshti University of Medical Sciences; Department of Epidemiology and Biostatistics, Tehran University of Medical Sciences, Tehran, Iran
5 School of Medicine, Shahid Beheshti University of Medical Sciences, Tehran, Iran

Date of Submission18-May-2014
Date of Acceptance21-Sep-2014
Date of Web Publication24-Nov-2015

Correspondence Address:
Hamideh Sabbaghi
Ophthalmic Research Center, Shahid Beheshti University of Medical Sciences, No. 23, Paidarfard St., Boostan 9 St., Pasdaran Ave., Tehran 16666
Iran
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/2008-322X.170360

Rights and Permissions
  Abstract 

Purpose: To determine the accuracy of photorefraction and autorefraction as compared to cycloautorefraction and to detect the repeatability of photorefraction. Methods: This diagnostic study included the right eyes of 86 children aged 7-12 years. Refractive status was measured using photorefraction (PlusoptiX SO4, GmbH, Nürnberg, Germany) and autorefraction (Topcon RM800, USA) with and without cycloplegia. Photorefraction for each eye was performed three times to assess repeatability. Results: The overall agreement between photorefraction and cycloautorefraction was over 81% for all refractive errors. Photorefractometry had acceptable sensitivity and specificity for myopia and astigmatism. There was no statistically significant difference considering myopia and astigmatism in all comparisons, while the difference was significant for hyperopia using both amblyogenic (P = 0.006) and nonamblyogenic criteria (P = 0.001). A myopic shift of 1.21 diopter (D) and 1.58 D occurred with photorefraction in nonamblyogenic and amblyogenic hyperopia, respectively. Using revised cut-off points of + 1.12 D and + 2.6 D instead of + 2.00 D and + 3.50 D improved the sensitivity of photorefractometry to 84.62% and 69.23%, respectively. The repeatability of photorefraction for measurement of myopia, astigmatism and hyperopia was acceptable (intra-cluster correlation [ICC]: 0.98, 0.94 and 0.77, respectively). Autorefraction results were significantly different from cycloautorefraction in hyperopia (P < 0.0001), but comparable in myopia and astigmatism. Also, noncycloglegic autorefraction results were similar to photorefraction in this study. Conclusion: Although photorefraction was accurate for measurement of myopia and astigmatism, its sensitivity for hyperopia was low which could be improved by considering revised cut-off points. Considering cut-off points, photorefraction can be used as a screening method.

Keywords: Accuracy; Autorefraction; Cycloautorefraction; Photorefraction; Repeatability


How to cite this article:
Rajavi Z, Sabbaghi H, Baghini AS, Yaseri M, Sheibani K, Norouzi G. Accuracy and repeatability of refractive error measurements by photorefractometry. J Ophthalmic Vis Res 2015;10:221-8

How to cite this URL:
Rajavi Z, Sabbaghi H, Baghini AS, Yaseri M, Sheibani K, Norouzi G. Accuracy and repeatability of refractive error measurements by photorefractometry. J Ophthalmic Vis Res [serial online] 2015 [cited 2019 Aug 25];10:221-8. Available from: http://www.jovr.org/text.asp?2015/10/3/221/170360




  Introduction Top


Refractive errors have been introduced as the second cause of treatable blindness worldwide.[1] The vision 2020 strategy of the World Health Organization (WHO) has also considered refractive errors as one of the leading causes of treatable blindness along with ocular disorders such as cataract, glaucoma, trachoma and onchocerciasis.[2] Uncorrected refractive errors are the most important factor predisposing to amblyopia among young children but can be controlled using accurate and sensitive screening.[3],[4]

Visual acuity (VA) assessment among young children is difficult and unreliable,[5],[6] therefore indirect investigation for amblyopia through measurement of refractive errors is advantageous.[7] Cycloplegic refraction is considered as the gold standard for detecting refractive errors,[3] although its application in screening programs entails problems such as being time consuming, requirement for supervision to monitor cycloplegia side effects, and increasing screening expenses and complications. In addition, it requires cooperation by children which is not easy to obtain in some children.[3],[5],[8],[9]

The photorefractometer is a non-contact, infrared video recorder [10] which has been introduced as a useful method for screening of amblyopia especially in preverbal children aged less than 4 years according to the American Association of Pediatric Ophthalmology and Strabismus (AAPOS).[11]

Early photorefractometers were not able to accurately detect astigmatism as compared to available autorefractors, however, modifications have been done to improve their accuracy since they were first introduced in 1974,[12] for instance, cylinder axis is determined in three meridians in newer models to improve assessment of astigmatism.[5] The newer generation of photorefractometer is also equipped with sound stimuli and improved software algorithms allowing faster data processing and statistical analysis, and also providing automatic image recording.[13]

One advantage of photorefractometry is that it can simultaneously detect small angle strabismus,[10] anisometropia, pupillary distance (PD), ptosis and media opacities in both eyes.[14] Furthermore, accommodation might be more relaxed since measurements can be performed at longer working distances (1 m or more) as compared to autorefractors.[15] Photorefractometers produce noises and flashing lights to help children fixate for a few seconds during measurements.[10] The other advantage of photorefractometers over autorefractors is that they are the method of choice for examination of young and uncooperative children and also mentally handicapped patients since subjects do not have to put their head and chin on a chin rest of autorefractometer.[15]

A wide range of sensitivity (63% to 94%) and specificity (62% to 99%) values for photorefractometry have been reported by Schimitzek and Haase.[3] Some studies have concluded that photorefraction has sufficient accuracy and reliability [9],[16] and could be considered as an alternative method for widespread school screening programs.[10],[11],[13] Schimitzek and Lagrèze also believed that photorefractometry is a useful method for ophthalmologists who are not skilled with retinoscopy or when refractive errors cannot be detected by autorefraction.[5] However, Ayse et al [15] found that the PlusoptiX SO4 is not an accurate means to estimate refraction in children.

The current study aimed to determine the accuracy and repeatability of PlusoptiX SO4 as compared to cycloautorefraction as the gold standard. In addition, autorefraction was compared with cycloautorefraction to detect the preferred method for screening refractive errors.


  Methods Top


This diagnostic study included 86 right eyes of 86 primary school children aged 7-12 years who were examined at the strabismus and pediatric clinic of Imam Hossein Medical Center, Tehran, Iran in 2013 to indicate the accuracy and repeatability of PlusoptiX SO4 (GmbH, Nürnberg, Germany). Children with mental retardation, ocular abnormalities, media opacity, impaired fixation, large angle of strabismus (more than 10 Δ), refractive errors exceeding the measurement range of the photorefractometer (−7.00 diopter [D] to + 5.00D), pupillary abnormalities and severe ptosis in whom photorefractometry could not exactly measure the refractive status, were excluded.

Prior to examination, the parents were interviewed by an optometrist to assess the child's systemic and ocular history, all details of the procedure were explained and written informed consent was obtained from all parents. This study was approved by the ethic committee of Ophthalmic Research Center, Shahid Beheshti University of Medical Sciences, Tehran, Iran.

Ptosis and nystagmus were detected by direct observation. VA was assessed at distance using Snellen E-chart with 5 letters on each line of a Yang vision tester (SIFI Diagnostic SPA, Via Castellana, Trevise, Italy) in a day light room with or without present glasses. First photorefraction was performed on all children at one meter distance by a trained operator while both eyes were open and repeated 3 times; then, refractive errors were rechecked by an autorefractometer (Topcon RM-8800; Topcon Medical, Oakland, NJ, USA) without cycloplegia. Finally, cycloautorefraction was performed 30-45 min after instillation of cyclopentolate 1% and tropicamide 1% eye drops at 5 min interval, as the gold standard method.

Emmetropia was considered as hyperopia (Sph)<+2.00D, myopia (Sph)>−0.50D and astigmatism <0 .75D. Also, nonamblyogenic refractive errors were considered as myopic Sph from -0.50 up to -3.00D and hyperopic Sph from +2.00 up to +3.50D and astigmatism 0.75 up to 1.50D.

Amblyogenic criteria were defined as myopic Sph ≤-3.00D, hyperopia Sph <+3.50D and astigmatism <1.50D according to AAPOS guidelines.[7]

With-the-rule astigmatism was considered if the axis was 180 ± 30°, while against-the-rule astigmatism was defined as axis within 90 ± 30°; oblique astigmatism was defined when the axis was between 30° and 60° or between 120° and 150°. Weighted axis difference was calculated by the following formula in which the axis difference of tested methods was weighted with the cylindrical (Cyl) power of the basic method in each comparison.[5],[10]

Weighted axis difference = 2 × Basic cylinder power sin (α2− α1)

α1 and α2 were the cylindrical axes of tested and basic methods in degrees, respectively.

Cycloautorefraction was considered as the gold standard during all comparisons, while autorefraction was defined as a basic method when auto and photorefraction were compared. Ocular deviation was evaluated by alternate cover test and the Krimsky method for children with VA <20/200 and VA <20/200, respectively. Ocular movements were also assessed in nine cardinal gazes to detect ocular muscle dysfunctions. The examinations were followed by direct ophthalmoscopy and slit lamp examination through a dilated pupil to explore pathological disorders in the anterior or posterior segments. Photorefraction for each child was performed three times by one operator while all conditions were kept stable for determining the repeatability of PlusoptiX SO4.

Statistical Analysis

To present data, mean values, standard deviations, medians and ranges were applied and repeated measure analyses were used to compare the results in multiple measurements. The repeatability of a set of measurements was evaluated using ICC (intra-cluster correlation) and the results of two sets were compared by paired sample t-test. To evaluate the diagnostic ability of photorefractometry, we used sensitivity, specificity, positive predictive value (PPV) and negative predictive values (NPV), as well as positive and negative likelihood ratio (LR) and receiver operating characteristic (ROC) curves. The correlations were evaluated using Pearson correlation. All statistical analyses were performed by SPSS software (version 21.0, IBM Co., Chicago, IL, USA).


  Results Top


The accuracy and repeatability of PlusoptiX SO4 were measured in 57 (66.3%) female and 29 (33.7%) male subjects. We compared PlusoptiX SO4 with cycloautorefraction as the gold standard regarding measurement of amblyogenic and nonamblyogenic refractive errors.

Comparison of Photorefraction and Cycloautorefraction

[Table 1] compares measurement of nonamblyogenic refractive errors using photorefraction and cycloautorefraction in 86 children. There was no difference between these two methods for measuring myopia and astigmatism, whereas a significant difference was observed in hyperopic measurements. There was 1.2D lower hyperopia (myopic shift) detected with photorefraction due to lack of cycloplegia (P = 0.001).
Table 1. Comparison of nonamblyogenic refractive error measurements between photorefraction and cycloautorefraction

Click here to view


[Table 2] compares measurement of amblyogenic refractive errors using photorefraction and cycloautorefraction. There was no statistically significant difference in the measurement of myopia and astigmatism while the difference for hyperopia was statistically significant (1.58 D, P = 0.006).
Table 2. Measurement of amblyogenic refractive errors comparing photorefraction and autorefraction with cycloautorefraction

Click here to view


Sensitivity, specificity and other coefficient indices of PlusoptiX SO4 are detailed in [Table 3]. The sensitivity of photorefraction ranged from 46.15% to 94.74% and its specificity was 88.14% to 100%. Positive and negative predictive values varied from 70.83% to 100% and from 85.25% to 96.1%, respectively. The overall agreement between the two methods was over 81% for all refractive error criteria. Kappa index was highest for astigmatism in the range of 0.75 up to 1.50D and gradually decreased for myopia; the smallest was calculated for hyperopia.
Table 3. Coefficient indices for photorefraction and suggested adjusted cutoff points as compared to cycloautorefraction

Click here to view


The largest difference between readings was observed in hyperopic patients as indicated by the largest distance from the cut-off point and the base criteria [Table 3].

The highest correlation between photorefraction and cycloautorefraction was observed for astigmatism (r = 0.946) whereas the weakest correlation was related to spherical equivalent (SE) of hyperopia (r = 0.588) according to Pearson correlation [Figure 1].
Figure 1. Pearson correlations between photorefraction and cycloautorefraction for hyperopia, myopia and astigmatism (r, 0.588; r, 0.915 and r, 0.946, respectively; P < 0.001 for all).

Click here to view


Receiver operating characteristics (ROC) curves were plotted to obtain the best cut-off points for refractive errors for amblyopia screening. The largest area under the curve belonged to myopia in the range of -0.50 up to -3.00D (n = 9) among nonamblyogenic refractive errors which reflects the greater ability of PlusoptiX SO4 in measuring myopia [Figure 2].
Figure 2. ROC curves for detecting nonamblyogenic refractive errors using photorefraction. Area under curve was 0.867 (95% CI: 0.805-0.930), 0.976 (95% CI: 0.939-0.995) and 0.954 (95% CI: 0.925-0.982) for hyperopia, myopia and cylindrical power, respectively.

Click here to view


[Figure 3] shows differences between the two methods in terms of SE, spherical and cylindrical power, and weighted axis. The maximum difference of photorefraction and cycloautorefraction was found for SE of ±1.00D, spherical refractive error of ±1.00D, cylindrical powers in the range of − 0.25D to + 0.50D and weighted axis of plano to +1.00D.
Figure 3. Difference of photorefraction and cycloautorefraction in measuring the spherical equivalent, sphere, cylindrical powers and weighted axis. SE, spherical equivalent; Sph, sphere; Cyl, cylinder.

Click here to view


Repeatability of PlusoptiX SO4

[Table 4] shows mean ± SD and median refractive errors based on three measurements in 86 children using PlusoptiX SO4. The calculated ICCs showed that PlusoptiX SO4 was repeatable for measurement of myopia, astigmatism and hyperopia (ICC of 0.98, 0.94 and 0.77, respectively).
Table 4. The repeatability of PlusoptiX SO4 for measurement of refractive errors

Click here to view


Comparison of Photorefraction and Autorefraction

There was no statistically significant difference between autorefraction and photorefraction for measurement of all refractive errors even hyperopia measurements.

Comparison of Autorefraction and Cycloautorefraction

[Table 2] shows a significant difference between autorefraction and cycloautorefraction for measurement of hyperopia among amblyogenic refractive errors; the underestimation by autorefraction was approximately 1.72D (P < 0.001).

[Figure 4] compares autorefraction and cycloautorefraction. The greatest difference in SE and spherical power was from − 0.50D to plano, and from − 1.00D to plano, respectively and the largest difference in cylindrical power was from plano to +0.25D. The weighted axis of noncycloplegic autorefraction differed from plano to +1.00D as compared to cycloplegic autorefraction measurements.
Figure 4. Difference of autorefraction and cycloautorefraction in measuring the spherical equivalent, sphere, cylindrical powers and weighted axis. SE, spherical equivalent; Sph, sphere; Cyl, cylinder.

Click here to view



  Discussion Top


Early visual screening plays an important role in decreasing the prevalence of amblyopia.[10],[17],[18] It is necessary to apply new methods with sufficient sensitivity and specificity for vision screening.[10] In the current study, three methods including photorefarction and autorefraction (with and without cycloplegia) were used to measure refractive errors. There was no statistically significant difference in myopia and astigmatism measurements in all methods, but hyperopia showed a significant difference between cycloautorefraction and each of photorefraction or autorefraction. By considering cut-off points, photorefraction can be also used as a screening tool [Table 1] and [Table 2].

We performed photorefraction without cycloplegia based on the manufacturer's recommendation, since induced peripheral aberrations due to dilated pupil makes measurement of astigmatism more difficult and there is a possibility of off-axis refraction.[5],[15] Erdurmus et al [8] reported that measuring refractive errors with dilated pupils (>8 mm) is difficult using PlusoptiX CRO3. Ayse et al [15] obtained inaccurate results with PlusoptiX SO4 under cycloplegia especially for cylindrical assessment. Schimitzek and Lagrèze [5] believed that cycloplegia may improve the accuracy of photorefraction for measuring SE, but decreases the accuracy of cylindrical power and axis. Moreover, using cycloplegic drops needs to be supervised for side effects which would increase screening expenses and complications.

To determine the preferable refractive screening method, we simultaneously compared photorefraction, autorefraction with cycloautorefraction (as the gold standard) and found no difference between photorefraction and noncycloautorefraction when comparing each of them with cycloautorefraction, therefore both of these methods can be used for refractive error screening.

The current study showed an overall agreement between photorefraction and cycloautorefraction ranging from 81.18% to 94.12% using both amblyogenic and non-amblyogenic refractive error criteria [Table 3]; which is in line with the results of other studies reporting 89.7%,[9] 84%[19] and 94%[13] agreements. There was a significant correlation between the two methods for measurement of myopia, and astigmatism [Figure 1] and [Figure 2] compatible with our previous study [9] and the report by Erdurmus et al.[8]

A myopic shift was observed with amblyogenic (−1.58 D) and nonamblyogenic (−1.21 D) hyperopia based on photorefractive readings as compared to cycloautorefraction [Table 1] and [Table 2] which is consistent with other studies.[5],[8],[9],[15],[16] This is not an unexpected finding, since noncycloplegic photorefraction at 1 m needs at least 1.00D of accommodation. The myopic shift in our study was close to 1.00D, while Erdurmus et al [8] found the amount to be 0.70D. This discrepancy may be attributed to different age ranges as well as a dissimilar number of hyperopic cases in the two studies. Schaeffel et al [16] found 2.40 D of myopic shift in spite of fogging the eyes using + 3.00D trial lenses. Although Arthur et al [13] found more myopic shift with higher hyperopia, this was not considerable in our study. Schimitzek and Lagrèze [5] also reported − 0.73 ± 1.25 D of myopic shift which was lower than our study due to the wide age range in their study (2-81 years old).

Based on our results, PlusoptiX SO4 had acceptable sensitivity and specificity for measuring myopia and astigmatism in amblyopia screening [Table 3]; however, it was necessary to modify hyperopic referral criteria to increase the sensitivity of the device for hyperopia. According to ROC curves the referral criteria were set as +1.12D and +2.60D instead of +2.00D and +3.50D for hyperopic patients, respectively. The sensitivity of PlusoptiX SO4 at these revised cutoff points can be improved from 65.38% to 84.62% and from 46.15% to 69.23%, respectively.

We found photorefraction to be an accurate and repeatable method for measurement of refractive errors particularly for myopia and astigmatism [Table 1],[Table 2],[Table 3],[Table 4] which is in agreement with the study by Allen et al [20] Considering the small size of myopic patients in the present study, further investigation in this regard seems necessary.

In summary, the PlusoptiX SO4 photorefractometer and autorefraction accurately measured myopic and cylindrical powers, however, their sensitivity for hyperopia was low as compared to cycloautorefraction as the gold standard. Hyperopic measurements will be improved if adjusted cutoff points are applied. Therefore, photorefraction can be used as a screening method with considering above mentioned revised cut-off points.

Acknowledgments

This study was supported by the Ophthalmic Research Center, Shahid Beheshti University of Medical Sciences and Basir Eye Safety Research Center.

The authors would like to thank all the children and their families who participated in this study, as well as Shadi Akbarian, Masoumeh Kalantarion, Simin Kalantarion, Bahar Kheiri, Shirin Mohebi and Bahar Safdari for their efforts in coordinating the fieldwork.

Source of Support:

Nil.

Conflict of Interest:

None declared.

 
  References Top

1.
Dandona R, Dandona L. Refractive error blindness. Bull World Health Organ 2001;79:237-243.  Back to cited text no. 1
    
2.
Foster A, Resnikoff S. The impact of Vision 2020 on global blindness. Eye (Lond) 2005;19:1133-1135.  Back to cited text no. 2
    
3.
Schimitzek T, Haase W. Efficiency of a video-autorefractometer used as a screening device for amblyogenic factors. Graefes Arch Clin Exp Ophthalmol 2002;240:710-716.  Back to cited text no. 3
    
4.
Salcido AA, Bradley J, Donahue SP. Predictive value of photoscreening and traditional screening of preschool children. J AAPOS 2005;9:114-120.  Back to cited text no. 4
    
5.
Schimitzek T, Lagrèze WA. Accuracy of a new photo-refractometer in young and adult patients. Graefes Arch Clin Exp Ophthalmol 2005;243:637-645.  Back to cited text no. 5
    
6.
Leman R, Armitage MD, Arnold RW. The receiver-operator curve for flip-card surround HOTV in younger school children: Validation of a simple flip card for school acuity testing. Am Orthopt J 2005;55:128-135.  Back to cited text no. 6
[PUBMED]    
7.
Donahue SP, Arnold RW, Ruben JB; AAPOS Vision Screening Committee. Preschool vision screening: What should we be detecting and how should we report it? Uniform guidelines for reporting results of preschool vision screening studies. J AAPOS 2003;7:314-316.  Back to cited text no. 7
    
8.
Erdurmus M, Yagci R, Karadag R, Durmus M. A comparison of photorefraction and retinoscopy in children. J AAPOS 2007;11:606-611.  Back to cited text no. 8
    
9.
Rajavi Z, Parsafar H, Ramezani A, Yaseri M. Is noncycloplegic photorefraction applicable for screening refractive amblyopia risk factors? J Ophthalmic Vis Res 2012;7:3-9.  Back to cited text no. 9
    
10.
Matta NS, Singman EL, Silbert DI. Performance of the Plusoptix vision screener for the detection of amblyopia risk factors in children. J AAPOS 2008;12:490-492.  Back to cited text no. 10
    
11.
Matta NS, Singman EL, Silbert DI. Performance of the plusoptiX S04 photoscreener for the detection of amblyopia risk factors in children aged 3 to 5. J AAPOS 2010;14:147-149.  Back to cited text no. 11
    
12.
Howland HC, Howland B. Photorefraction: a technique for study of refractive state at a distance. J Opt Soc Am 1974;64:240-249.  Back to cited text no. 12
[PUBMED]    
13.
Arthur BW, Riyaz R, Rodriguez S, Wong J. Field testing of the plusoptiX S04 photoscreener. J AAPOS 2009;13:51-57.  Back to cited text no. 13
    
14.
Miller JM, Dobson V, Harvey EM, Sherrill DL. Cost-efficient vision screening for astigmatism in native American preschool children. Invest Ophthalmol Vis Sci 2003;44:3756-3763.  Back to cited text no. 14
    
15.
Ayse YK, Onder U, Suheyla K. Accuracy of Plusoptix S04 in children and teens. Can J Ophthalmol 2011;46:153-157.  Back to cited text no. 15
    
16.
Schaeffel F, Mathis U, Brüggemann G. Noncycloplegic photorefractive screening in pre-school children with the "PowerRefractor" in a pediatric practice. Optom Vis Sci 2007;84:630-639.  Back to cited text no. 16
    
17.
Kvarnström G, Jakobsson P, Lennerstrand G. Visual screening of Swedish children: an ophthalmological evaluation. Acta Ophthalmol Scand 2001;79:240-244.  Back to cited text no. 17
    
18.
Kvarnström G, Jakobsson P, Lennerstrand G, Dahlgaard J. Preventable vision loss in children: A public health concern? Am Orthopt J 2006;56:3-6.  Back to cited text no. 18
    
19.
Clausen MM, Arnold RW. Pediatric eye/vision screening. Referral criteria for the pedia vision plus optix s 04 photoscreener compared to visual acuity and digital photoscreening. Kindergarten computer photoscreening. Binocul Vis Strabismus Q 2007;22:83-89.  Back to cited text no. 19
    
20.
Allen PM, Radhakrishnan H, O'Leary DJ. Repeatability and validity of the PowerRefractor and the Nidek AR600-A in an adult population with healthy eyes. Optom Vis Sci 2003;80:245-251.  Back to cited text no. 20
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4]
 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4]


This article has been cited by
1 The Repeatability of Values Measured Using the Spot Vision Screener in Healthy Children and Children with Refractive Errors
Tsukasa Satou,Tsuyoshi Nogami,Yoshiaki Takahashi,Misae Ito,Takahiro Niida
Journal of Binocular Vision and Ocular Motility. 2019; : 1
[Pubmed] | [DOI]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Introduction
Methods
Results
Discussion
References
Article Figures
Article Tables

 Article Access Statistics
    Viewed1362    
    Printed15    
    Emailed3    
    PDF Downloaded179    
    Comments [Add]    
    Cited by others 1    

Recommend this journal


[TAG2]
[TAG3]
[TAG4]