REVIEW ARTICLE
Year : 2015  |  Volume : 10  |  Issue : 4  |  Page : 461-468

Descemet's stripping automated endothelial keratoplasty tissue insertion devices


1 Department of Mechanical Engineering, The University of Texas at Arlington, Arlington, TX, USA
2 Department of Ophthalmology, UT Sowthwestern Medical Center, Dallas, TX, USA

Date of Submission04-Feb-2015
Date of Acceptance28-Feb-2015
Date of Web Publication18-Feb-2016

Correspondence Address:
Salman Nasir Khan
Box 19023, 500 W, First St. 204, Woolf Hall, Arlington, TX
USA
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/2008-322X.176899

Rights and Permissions
  Abstract 

This review study provides information regarding the construction, design, and use of six commercially available endothelial allograft insertion devices applied for Descemet's stripping automated endothelial keratoplasty (DSAEK). We also highlight issues being faced in DSAEK and discuss the methods through which medical devices such as corneal inserters may alleviate these issues. Inserter selection is of high importance in the DSAEK procedure since overcoming the learning curve associated with the use of an insertion device is a time and energy consuming process. In the present review, allograft insertion devices were compared in terms of design, construction material, insertion technique, dimensions, incision requirements and endothelial cell loss to show their relative merits and capabilities based on available data in the literature. Moreover, the advantages/disadvantages of various insertion devices used for allograft insertion in DSAEK are reviewed and compared. The information presented in this review can be utilized for better selection of an insertion device for DSAEK.

Keywords: Allograft Inserter; Allograft Insertion Device; Corneal Injectors; Descemet's Stripping Automated Endothelial Keratoplasty; Descemet's Stripping Endothelial Keratoplasty


How to cite this article:
Khan SN, Shiakolas PS, Mootha VV. Descemet's stripping automated endothelial keratoplasty tissue insertion devices. J Ophthalmic Vis Res 2015;10:461-8

How to cite this URL:
Khan SN, Shiakolas PS, Mootha VV. Descemet's stripping automated endothelial keratoplasty tissue insertion devices. J Ophthalmic Vis Res [serial online] 2015 [cited 2017 Sep 24];10:461-8. Available from: http://www.jovr.org/text.asp?2015/10/4/461/176899


  Introduction Top


The cornea is the outermost transparent part of the eye which covers the iris, pupil and anterior chamber (AC). The cornea is transparent and comprises of five layers, namely, epithelium, Bowman's membrane, stroma, Descemet's membrane and endothelium.[1],[2] The endothelium, the innermost layer, is an extremely thin layer responsible for keeping the cornea in its dehydrated state, which is vital for maintaining corneal transparency.[3],[4] The endothelial cells (ECs) act as small pumps, which suck fluid from the corneal stroma and pump it back to the AC hence preserving corneal transparency. These cells regenerate at an extremely slow rate as compared to cell decay, since they are arrested in the G1 cell cycle.[5],[6] Therefore cell counts rapidly decrease until the age of 25 after which the rate of decay slows down.[7],[8] For clear vision, the EC count should stay above a certain critical amount; however diseases such as Fuchs' dystrophy, pseudophakic or aphakic bullous keratopathy, posterior polymorphous dystrophy, iridocorneal endothelial syndrome, endothelial decompensation and failed penetrating keratoplasty (PK), result in a decrease in EC count and eventually, an endothelial layer transplant becomes necessary.[9] Numerous surgeries have been developed in the past for endothelial layer transplant such as PK, deep lamellar endothelial keratoplasty, Descemet's stripping automated endothelial keratoplasty (DSAEK) and Descemet's membrane endothelial keratoplasty. PK and DSAEK are the most popular choices among surgeons.[10],[11]

In 2013, the number of corneal tissues used for transplant/keratoplasty in the US alone was 72,736.[12] This number was largely shared between PK (36,998) and endothelial keratoplasty (EK) (27,298) procedures with with larger share for PK. However, EK is quickly improving showing approximately 12% increase from previous years, and this number is expected to increase every year along with improvements in health care and life expectancy.[10],[13] DSAEK is a type of keratoplasty that has become the gold standard in EK in recent years. This procedure is effective, fairly reliable and shows low rates of complications, however, it is evolving towards achievement of more acceptable results.[14]

With DSAEK the diseased endothelium and Descemet's membrane are replaced with an excised posterior allograft approximately 130-240 μm in thickness. This allograft comprises of posterior corneal stroma, healthy Descemet's membrane and endothelium. This donor allograft once trephined, is inserted into the AC through a side incision created in the cornea, limbus or sclera. The advantages of EK are small incision size, a structurally stronger cornea, requirement of few to no sutures, reduced operative time, lower risk of allograft rejection, cost effectiveness, stability of refraction, and swift visual rehabilitation [15],[16],[17],[18],[19],[20],[21],[22],[23],[24],[25] However, the demerits of DSAEK are a significant initial loss in endothelial cells density (ECD) as well as AC collapse, intraoperative ECD loss particularly in Asian patients with shallow ACs and in eyes with high vitreous pressures, and ECD reduction during microkeratome cutting of donor allografts, handling of corneal allograft and air bubble insertion into the AC.[26],[27] The major ECD loss occurs during allograft insertion and subsequent unfolding in the AC,[9] which can potentially lead to problems such as primary allograft failure and corneal opacity.[22],[27],[28],[29] AC collapse depends on incision size, allograft insertion mechanism, and AC maintainer.[30] The majority of issues affecting DSAEK outcomes stem from the allograft insertion technique. In this regard, great attention has recently been directed toward developing allograft insertion devices to minimize trauma to the allograft tissue during insertion and unfolding, hence diminishing EC loss and AC collapse.

This review aims to compare the designs of various corneal inserters and discusses the performance of these insertion devices by reviewing their construction/design and the results of related clinical studies. Performance factors are considered taking into account ECD loss percentage, AC collapse, ease of use, number of hands required by the surgeon for use, required incision size, reusability and cost. Clinical studies conducted on the injectors are also compared. Although these studies were performed by different surgeons, the comparison is still valid as ECD loss or successful surgery is more dependent on technique and devices rather than surgeon's expertise.[31] Another factor of variability was the difference in storage times of the donor allografts which are utilized in different clinical studies. However, a recent study proved that allograft storage time was of little consequence in terms of ECD loss or the success of the DSAEK surgery.[32] This review will provide helpful information regarding the current market of DSAEK inserter for surgeons and medical device professionals.


  Endothelial Allograft Insertion Devices Top


Forceps have traditionally been utilized for allograft insertion; however, corneal inserters seem a necessity because corneal allografts inserted using forceps experience more trauma resulting in greater ECD loss. All available inserters have been designed with the primary objective to protect the allograft from folding and reduce incision compression pressure (ICP). Almost all inserters have shown better ECD results in clinical trials as compared to forceps.[33],[34],[35],[36],[37],[38] Allografts inserted with forceps display a definite pattern of trauma such as parallel bands and orthogonal wrinkles due to folding and compression of the tissue during insertion through the incision.[22] The initial trauma experienced by the allograft is even greater in Asian patients as they have shallow ACs, making allograft insertion and unfolding more challenging.[39] Reduction in initial endothelial allograft trauma was mainly considered while developing a secure method in which the allograft is protected from ICP and folding effects. The developed insertion devices may broadly be categorized into two groups based on their injection mechanism; namely pull through designs (glides) and push in designs (injectors). Most of these devices require an incision size in the range of 3-5 mm to inject a lamellar allograft with a diameter >8 mm and thickness of 130-240 μm.[10] This paper will compare the design of three glides and three injectors which are commercially available and extensively used. The glides discussed are Endoglide (Angiotech, Vancouver, BC, Canada), Macaluso inserter (E. Janach, Como, Italy) and Busin glide (Asico, Westmont, IL, USA). The reviewed injectors include Endoserter (Ocular Systems, Winston-Salem, NC, USA), Endoshield/Endoinjector (Keramed, San Jose, CA, USA), and Neusidl injector (Fischer Surgical, Arnold, MO, USA) [Table 1].
Table 1. Characteristics of corneal inserters

Click here to view



  Endothelial Glides Top


To operate endothelial glides, a corneal allograft is pulled inside the AC from the far side using forceps. This is a two hand technique with the inherent advantage of correct allograft unfolding and orientation inside the AC. The allograft undergoes considerably less trauma during insertion as compared with insertion using forceps. Allograft displacement, however, may occur due to AC maintainer interference. Conditions with high intraocular pressure may lead to AC collapse, complicating graft handling since glides require the use of both hands by the surgeon. Endothelial glides come in different shapes and designs depending on the manufacturer and some of these glides have already received Food and Drug Administration (FDA) clearance.


  Busin Glide Top


The Busin glide is a 126 mm long all metallic, FDA approved, reusable corneal allograft insertion device which was invented by Kobayashi and Busin. It consists of a handle with a pan at the end. The allograft is placed endothelial side up inside the marked space of the pan. This marked space is 8 mm in diameter, and it aids in proper centering of the allograft. Once the allograft is in correct position, grasping forceps are used to bring the allograft towards the peripheral opening of the device. While proceeding towards the opening, the allograft rolls and takes the shape of the peripheral tip of the device. Having the allograft at the tip, aids in grasping it once the tip of the Busin glide is inside the AC. The nontapered/elliptical peripheral opening of the Busin glide is inserted into the AC through a 5.0 mm incision in the eye [Figure 1].[40] It has a semicircular cutout, which aids in grasping the allograft once inside the AC. Note that during this procedure AC pressure is maintained via an AC maintainer. The elliptical tip of the Busin glide protects the allograft from ICP and prevents folding of the allograft as well. The maximum diameter of the allograft that Busin glide can hold without overlap is 8 mm. The majority of surgeons place the Busin glide slightly outside the incision, to avoid AC collapse; however, this exposes the allograft to ICP and leads to endothelial trauma. Other surgeons insert the glide inside the incision to save the allograft from ICP; nevertheless, this increases the possibility of AC collapse, which in turn may lead to other complications.[41] The performance of this device is well documented along with ECD loss ranging from 20 to 47.5% six months after the operation.[33],[34],[42],[43]
Figure 1. Busin glide inserter

Click here to view



  Macaluso Inserter Top


The Macaluso inserter is an all metallic reusable glide which is named after its inventor Claudio Macaluso [Figure 2].[44],[45],[46] This device is being used in Europe and has not yet been approved by the FDA for use in the USA. The Macaluso inserter employs a close chamber technique, which aids in maintaining a well formed AC.[45] The allograft is placed endothelial side up on the concave winglet, and some viscoelastic is applied, the winglet has a circular region marked for assisting in centration of the allograft. The allograft is then grasped with forceps and pulled towards the opening of the insertion device. While proceeding towards the tip, the allograft rolls and conforms to the shape of the device tip. Once the allograft is in a position at the tip, a plunger is used to seal the allograft inside the tip and consequently the allograft and cornea medium are confined in a closed space. The peripheral opening of the device has a tapered elliptical shape with a 'V' shaped cutout to provide easier access for 23 gauge coaxial grasping forceps. A 4.2 mm incision is usually recommended when using this device in DSAEK procedures. Furthermore, at all times during insertion, an AC maintainer is employed to maintain the AC depth. The maximum diameter of the donor allograft that can be held inside the Macaluso inserter is 8-8.5 mm according to the manufacturer. However, what sets this inserter apart from other glides is the provision of a push plunger which seals the system from the environment. Hence, when the device is inserted inside the AC for allograft injection, the AC and inserter act as one sealed system, allowing the AC to remain stable. The metallic body of the Macaluso inserter makes it non-transparent. However, it can be sterilized and reused. There is not a lot of information available in the published literature discussing the clinical performance of this device. More research is recommended to assess the postoperative outcomes of this device.
Figure 2. Macaluso endothelial inserter.[46]

Click here to view



  Tan Endoglide Top


The Tan Endoglide is an FDA approved, transparent, disposable endothelial allograft inserter invented by Donald Tan. It consists of a preparation base, glide cartridge and glide introducer [Figure 3]a.[37] The allograft is placed endothelial side up on the Endoglide cartridge, which is first slotted into the preparation base, and some lubricating liquid is applied. Grasping forceps are introduced from the tip of cartridge, and the endothelial allograft is pulled inside the cartridge. As the allograft is pulled towards the tip of cartridge, it takes the shape of the tip which induces a “double coil” configuration shape to the endothelial allograft [Figure 3]b. Once the allograft is at the tip, the Endoglide introducer is connected to the back of cartridge, sealing the back end in the process. The Endoglide is then flipped and inserted inside the AC through a 4.0-4.5 mm scleral or corneal incision,[47] with the AC maintainer already inside the AC. Once the Endoglide is inside the AC, it seals it and creates a closed system. It prevents fluid from flowing out through the incision or the glide itself, and once the glide is in place, grasping forceps are introduced inside the AC from the opposite side and the allograft is gently pulled inside the AC. The tip of the Endoglide is oval in shape and can hold an allograft up to 250 μm in thickness. However, what sets Endoglide apart is its “double coil” tip shape, which enables this device to use the space more efficiently hence it enables the insertion of a larger diameter allograft through a relatively smaller incision. Various DSAEK clinical studies have been conducted using the Endoglide for allograft insertion, and most of them show less ECD loss as compared to forceps insertion. ECD loss has been reported to range from 13 to 26% six months postoperatively.[35],[37],[41]
Figure 3. (a) Tan endoglide (b) Double coil.

Click here to view



  Endothelial Allograft Injectors Top


To operate endothelial injectors, a corneal allograft is inserted into the AC using a pushing mechanism. These are usually single hand operated devices in which the allograft endures considerably less trauma during insertion as compared to forceps insertion.[34] Some injectors have the added advantage of an inbuilt capability to maintain AC pressure hence an external AC maintainer is not required. The injector body is usually filled or coated with lubricating liquid before the allograft is inserted into the device. The inserted allograft takes the shape of the injector tip as it slides toward the opening. The lubricating liquid helps reduce friction and protects the allograft from damage. Injectors come in different shapes and designs depending on the manufacturer; some of these injectors have already received FDA clearance while others are still waiting.


  Neusidl Corneal Inserter Top


This device is named after its inventor William B Neusidl. It is an irrigation incorporated, nontransparent, disposable corneal inserter which is FDA approved [Figure 4].[48] It consists of a body with a flexible platform upfront where the allograft is placed endothelial side up, once the surface has been lubricated.[49] The flexible platform is able to hold 8-8.5 mm allografts and ensures that the delivery of the allograft is in the correct orientation; therefore the need for marking the stromal side of the allograft is eliminated. A lever is used to extend the flexible surface, the allograft is placed on this surface and subsequently the lever is used to retract the flexible surface along with the allograft inside the injector. While moving inside the tapered elliptical opening of the device, the allograft conforms to the shape of the inner wall of the elliptical tip and at all times it avoids touching or folding the endothelium. Once the allograft is securely retracted inside, the irrigation is attached and tested. The injector is then rotated such that it is oriented bevel down. In this orientation, the injector is inserted inside the AC through a 5.0-5.50 mm incision.[36] In fact, the bevel down orientation aids in unfolding of the allograft inside the AC.
Figure 4. Neusidl injector

Click here to view


A surgical stop is provided in the device, which results in better sealing of the wound and acts as an indicator for the surgeon confirming that the recommended insertion has been achieved inside the AC. This, however, is a slight generalization since AC dimensions of patients vary slightly. During the insertion of the allograft, the elliptical tip of the device protects the allograft from ICP, and the irrigation pressure ensures a deep chamber. Having a deep chamber is essential since it provides more space for the allograft to unfold which in turn means that the allograft will experience less trauma.[50] The irrigation provision in the Neusidl inserter is compatible with standard cataract surgery irrigation equipment, hence an AC maintainer is not required. Various DSAEK clinical studies have been conducted using the Neusidl inserter for allograft insertion and most of them show comparable or reduced ECD loss as compared to insertion forceps. ECD loss six 6 months postoperatively ranges from 13 to 33%.[51] However, some surgeons have highlighted that the allograft occasionally sticks to the platform and requires shaking of the device to get it off.[36] A lubricant to alleviate this problem would be beneficial.


  Endoinjector Top


Previously known as the Endoshield, the Endoinjector was invented by Yichieh Shiuey. It is a nontransparent, single use corneal allograft inserter, which has the European CE mark approval and is awaiting FDA approval.[52] It consists of an injector body, injector cartridge, and a foam tip plunger [Figure 5]. The injector cartridge comprises of a central ridge in the cartridge lumen, which guides the allograft to form a double coil as the allograft is advanced, therefore preventing allograft rotation in the process. The balanced salt solution (BSS) soaked allograft is placed endothelial side up inside the loading area, and fluid propulsion is used to guide the allograft forward to the beveled cartridge tip. The Endoinjector has been designed to atraumatically deliver 8-9 mm allografts inside the AC in the correct orientation.[52] Once the allograft is placed in the loading area and the wings of the cartridge are closed, the cartridge is loaded in the injector body and the wings are rotated clockwise for a right handed surgeon and vice versa. However, before the cartridge is loaded onto the injector body, Healon and BSS are used to plug and fill the cartridge in order to make the device act similar to a fluid chamber, thus making it a multi-step process. Once the cartridge is ready and loaded onto the injector body, pressure is applied to the thumb rest to move the allograft forward towards the end of the cartridge. Once the allograft is at the tip of the cartridge, an AC maintainer is inserted inside the AC. The Endoinjector is then rotated bevel down and inserted inside the AC through a 3.2 mm corneal incision, this bevel orientation aids in unfolding the allograft inside the AC. During insertion of the allograft, the Endoinjector protects the allograft from ICP, and the AC maintainer ensures a deep chamber. There are few clinical studies utilizing the Endoinjector for corneal allograft insertion in DSAEK. However, one study found the ECD loss to be 13%.[52] More results of clinical trials seem necessary to be published in the literature using this device for accurate assessment of postoperative outcomes of this device.
Figure 5. Endoinjector

Click here to view



  Endoserter Top


The Endoserter is an endothelial allograft insertion device invented by Keith Walter. It is an irrigation incorporated, nontransparent, disposable corneal inserter which is FDA approved [Figure 6].[53] It consists of a flexible platform where the allograft is placed endothelial side up, once the device has been flushed with BSS by connecting the irrigation tubing to the lower end connector of the Endoserter.[54] This flexible platform can accommodate allografts 8.5 mm or less in diameter and 175 μm in central thickness. Besides, it ensures the delivery of the allograft is in the correct orientation. A rotating wheel is used to retract the allograft carrier inside the injector. While moving inside the tip of the injector, the allograft conforms to the shape of the inner wall of the beveled tip and at all times touching or folding the endothelium is avoided. The injector is then rotated bevel down and in this orientation the Endoserter is inserted into the AC through a 4.0 mm incision. This bevel down orientation aids in unfolding the allograft inside the AC. During insertion of the allograft, the Endoserter protects the allograft from ICP, and the irrigation pressure ensures a deep chamber. The irrigation provision in the Endoserter is compatible with standard cataract irrigation equipment. A recent DSAEK clinical study conducted using the Endoserter reported a mean ECD loss of 29% six months postoperatively.[36] More results of clinical studies/trials need to be published in the open literature using this device so that its postoperative outcomes can be assessed.
Figure 6. Endoserter

Click here to view



  Discussion Top


DSAEK is an evolving procedure which has already demonstrated many advantages over traditional PK and is gradually becoming the procedure of choice among surgeons. Some challenges regarding its implementation still exist, which are being addressed to some extent with the use of endothelial inserters, since forceps have inherent limitations. Using forceps is a simple technique; however, forceps apply excessive pressure at the points of contact and do not protect the allograft from ICP, leading to higher postoperative ECD loss. Through the present review, it can be safely concluded that each of the six reviewed inserters has its own advantages and limitations regarding ECD loss, the probability of AC collapse, ease of use, number of hands required by the surgeon for usage, required incision size, reusability and cost [Table 1]. However, the primary objective of every inserter is to reduce the amount of trauma experienced by the allograft in order to reduce ECD loss sustained during the surgical procedure. All corneal inserters discussed in this review showed better postoperative ECD results as compared to traditional forceps. Endoglide, Endoinjector and Neusidl corneal inserters seemed to show better ECD results than all three injectors registering a minimum of 13% ECD loss according to published data.

Apart from ECD loss, another important issue faced by surgeons is flipping of the donor allograft inside the AC. Since visibility is low inside the AC, flipping of donor allograft poses a significant complication which may result in failure of the DSAEK procedure. All the six inserters discussed herein address this issue properly; however, since grasping forceps are used to insert donor allografts in glides, therefore glides have an almost perfect chance of donor allograft unfolding in the correct orientation. Some surgeons write, “S” or “Z” on the stromal side of the donor allograft so that they can unfold the allograft in the correct orientation in the AC. Nevertheless, the writing ink contains alcohol, which may seep through the stroma and damage ECs.[55] Thus, it would be best to avoid it altogether or the surgeon should write the letter on his glove or use a stamp, wait till the alcohol evaporates and then use it as a branding stamp to mark the donor allograft.[56] The postoperative ECD comparison results presented in this review provide a valuable overview of commercially available DSAEK inserters and present a much required perspective regarding DSAEK allograft inserters. However, current ECD results were acquired from clinical studies which involved patients from different demographics and utilized different mechanisms to measure postoperative ECD such as specular microscopy and vital dye staining with use of graphic software. Consequently, the comparisons made between the inserters lack a common denominator. To accurately compare ECD loss efficiency with these inserters, all six of them should be used in a single study in the future. At present, most clinical studies have not yet involved more than two DSAEK allograft inserters therefore a study employing and comparing all commercially available inserters seems necessary. Such a study will be able to act as a benchmark regarding inserter comparison and will aid in identification of the inserter that imparts the least trauma to the allograft. Furthermore, trauma patterns can be studied and correlated with allograft insertion parameters such as orientation angle and insertion speed, and hence techniques can be refined and standardized.

Apart from prevention of ECD loss and correct allograft unfolding, other factors of note in DSAEK are AC maintainer requirement and incision size. Both of these factors are of significance and to some extent correlated since a bigger incision leads to hypotony, invariably leading to AC collapse and allograft failure. A smaller incision reduces the probability of AC collapse and requires less sutures thereby reducing surgery time. Among the inserters reviewed herein, the Endoinjector requires the smallest incision size of 3.2 mm. All of the inserters in the present paper require an AC maintainer, except the Neusidl inserter and Endoserter. An inserter which does not require an AC maintainer is a definite advantage; however, a smaller incision size may be considered as a two-edged sword since on one hand it means less or no sutures and less chance of AC collapse, while on the other hand it entails more ECD loss.[30],[43] An optimum inserter head size/incision size needs to be investigated and studied. This knowledge will improve DSAEK outcomes, make it more streamlined and hence easier for inexperienced surgeons to adopt.

In summary, DSAEK is a relatively new and evolving surgical procedure with a respectable success rate. However, there is still room for improvement to achieve more ideal postoperative visual outcomes. Hence, corneal inserters, as compared to forceps, should be adopted since they seem to be promising for preventing complications, improving surgical outcomes and reducing the learning curve associated with the surgery. The inserters reviewed in this paper have their own advantages and limitations, therefore further studies/research on their efficacy should be performed. The results or feedback from these studies may assist to improve the designs of existing inserters and develop new inserters with the capability to address all prevailing issues.


  Acknowledgments Top


The authors would like to acknowledge that the research work presented in this paper was not funded by any organization.

 
  References Top

1.
Smerdon D. Anatomy of the eye and orbit. Curr Anaesth Crit Care 2000;11:286-292.  Back to cited text no. 1
    
2.
Khan S, Shiakolas PS, Rizwan M. An overview on performance characteristics of laser in-situ keratomileusis using lasers and identification of challenges. Micro Nanosys 2012;4:284-95.  Back to cited text no. 2
    
3.
Tuft SJ, Coster DJ. The corneal endothelium. Eye (Lond) 1990;4 (Pt 3):389-424.  Back to cited text no. 3
    
4.
Geroski DH, Matsuda M, Yee RW, Edelhauser HF. Pump function of the human corneal endothelium. Effects of age and cornea guttata. Ophthalmology 1985;92:759-763.  Back to cited text no. 4
[PUBMED]    
5.
Joyce NC. Proliferative capacity of the corneal endothelium. Prog Retin Eye Res 2003;22:359-389.  Back to cited text no. 5
    
6.
Carlson KH, Bourne WM, McLaren JW, Brubaker RF. Variations in human corneal endothelial cell morphology and permeability to fluorescein with age. Exp Eye Res 1988;47:27-41.  Back to cited text no. 6
    
7.
Laule A, Cable MK, Hoffman CE, Hanna C. Endothelial cell population changes of human cornea during life. Arch Ophthalmol 1978;96:2031-2035.  Back to cited text no. 7
[PUBMED]    
8.
Wilson RS, Roper-Hall MJ. Effect of age on the endothelial cell count in the normal eye. Br J Ophthalmol 1982;66:513-515.  Back to cited text no. 8
[PUBMED]    
9.
Price MO, Price FW Jr. Endothelial cell loss after descemet stripping with endothelial keratoplasty influencing factors and 2-year trend. Ophthalmology 2008;115:857-865.  Back to cited text no. 9
    
10.
Anshu A, Price MO, Tan DT, Price FW Jr. Endothelial keratoplasty: A revolution in evolution. Surv Ophthalmol 2012;57:236-252.  Back to cited text no. 10
    
11.
Eye bank Association of America. 2013 Eye Banking Statistical Report Washington DC USA. Retrieved from http://www.restoresight.org/wp-content/uploads/2014/04/2013_Statistical_Report-FINAL.pdf. 2013. p. 114.  Back to cited text no. 11
    
12.
Eye Bank Association of America. 2013 Eye Banking Statistical Report. Washington, DC, USA: Eye Bank Association of America; 2013.  Back to cited text no. 12
    
13.
Boimer C, Lee K, Sharpen L, Mashour RS, Slomovic AR. Evolving surgical techniques of and indications for corneal transplantation in Ontario from 2000 to 2009. Can J Ophthalmol 2011;46:360-366.  Back to cited text no. 13
    
14.
Lee JK, Hwang AM. Descemetâ s Stripping Automated Endothelial Keratoplasty â A Review. Eur Ophthalmic Rev 2009;3:71-3.  Back to cited text no. 14
    
15.
Clements JL, Bouchard CS, Lee WB, Dunn SP, Mannis MJ, Reidy JJ, et al. Retrospective review of graft dislocation rate associated with Descemet stripping automated endothelial keratoplasty after primary failed penetrating keratoplasty. Cornea 2011;30:414-418.  Back to cited text no. 15
    
16.
Thompson RW Jr, Price MO, Bowers PJ, Price FW Jr. Long-term graft survival after penetrating keratoplasty. Ophthalmology 2003;110:1396-1402.  Back to cited text no. 16
    
17.
Afshari NA, Gorovoy MS, Yoo SH, Kim T, Carlson AN, Rosenwasser GO, et al. Dislocation of the donor graft to the posterior segment in Descemet stripping automated endothelial keratoplasty. Am J Ophthalmol 2012;153:638-642.  Back to cited text no. 17
    
18.
Price MO, Fairchild KM, Price DA, Price FW Jr. Descemet's stripping endothelial keratoplasty five-year graft survival and endothelial cell loss. Ophthalmology 2011;118:725-729.  Back to cited text no. 18
    
19.
Chen ES, Terry MA, Shamie N, Hoar KL, Friend DJ. Descemet-stripping automated endothelial keratoplasty: Six-month results in a prospective study of 100 eyes. Cornea 2008;27:514-520.  Back to cited text no. 19
    
20.
Bahar I, Kaiserman I, Levinger E, Sansanayudh W, Slomovic AR, Rootman DS. Retrospective contralateral study comparing descemet stripping automated endothelial keratoplasty with penetrating keratoplasty. Cornea 2009;28:485-488.  Back to cited text no. 20
    
21.
Koenig SB, Covert DJ. Early results of small-incision Descemet's stripping and automated endothelial keratoplasty. Ophthalmology 2007;114:221-226.  Back to cited text no. 21
    
22.
Ide T, Yoo SH, Goldman JM, Perez V, O'Brien TP. Descemet-stripping automated endothelial keratoplasty: Effect of inserting forceps on DSAEK donor tissue viability by using an in vitro delivery model and vital dye assay. Cornea 2007;26:1079-1081.  Back to cited text no. 22
    
23.
Ang M, Mehta JS, Lim F, Bose S, Htoon HM, Tan D. Endothelial cell loss and graft survival after Descemet's stripping automated endothelial keratoplasty and penetrating keratoplasty. Ophthalmology 2012;119:2239-2244.  Back to cited text no. 23
    
24.
Bose S, Ang M, Mehta JS, Tan DT, Finkelstein E. Cost-effectiveness of Descemet's stripping endothelial keratoplasty versus penetrating keratoplasty. Ophthalmology 2013;120:464-470.  Back to cited text no. 24
    
25.
Prabhu SS, Kaakeh R, Sugar A, Smith DG, Shtein RM. Comparative cost-effectiveness analysis of descemet stripping automated endothelial keratoplasty versus penetrating keratoplasty in the United States. Am J Ophthalmol 2012;155:45-53.  Back to cited text no. 25
    
26.
Terry MA, Saad HA, Shamie N, Shah AK. Peripheral endothelial cell damage after trephination of donor tissue. Cornea 2009;28:1149-1152.  Back to cited text no. 26
    
27.
Hong A, Caldwell MC, Kuo AN, Afshari NA. Air bubble-associated endothelial trauma in descemet stripping automated endothelial keratoplasty. Am J Ophthalmol 2009;148:256-259.  Back to cited text no. 27
    
28.
Price FW Jr, Price MO. Descemet's stripping with endothelial keratoplasty in 200 eyes: Early challenges and techniques to enhance donor adherence. J Cataract Refract Surg 2006;32:411-418.  Back to cited text no. 28
    
29.
Alkatan H, Al-Rajhi A, Al-Shehri A, Khairi A. Histopathological findings of failed grafts following Descemet's stripping automated endothelial keratoplasty (DSAEK). Saudi J Ophthalmol 2012;26:79-85.  Back to cited text no. 29
    
30.
Ide T. Descemet's stripping automated endothelial keratoplasty injecting device. ExpertRev Ophthalmol 2009;4:5-9.  Back to cited text no. 30
    
31.
Chen ES, Terry MA, Shamie N, Hoar KL, Phillips PM, Friend DJ. Endothelial keratoplasty: Vision, endothelial survival, and complications in a comparative case series of fellows vs attending surgeons. Am J Ophthalmol 2009;148:26-31.  Back to cited text no. 31
    
32.
Terry MA, Shamie N, Straiko MD, Friend DJ, Davis-Boozer D. Endothelial keratoplasty: The relationship between donor tissue storage time and donor endothelial survival. Ophthalmology 2011;118:36-40.  Back to cited text no. 32
    
33.
Busin M, Bhatt PR, Scorcia V. A modified technique for descemet membrane stripping automated endothelial keratoplasty to minimize endothelial cell loss. Arch Ophthalmol 2008;126:1133-1137.  Back to cited text no. 33
    
34.
Bahar I, Kaiserman I, Sansanayudh W, Levinger E, Rootman DS. Busin guide vs forceps for the insertion of the donor lenticule in descemet stripping automated endothelial keratoplasty. Am J Ophthalmol 2009;147:220-226.e1.  Back to cited text no. 34
    
35.
Khor WB, Mehta JS, Tan DT. Descemet stripping automated endothelial keratoplasty with a graft insertion device: Surgical technique and early clinical results. Am J Ophthalmol 2011;151:223-232.  Back to cited text no. 35
    
36.
Griffin NB, Walter KA. A safe and convenient method of inserting and controlling donor tissue during endothelial keratoplasty. US Ophthal Rev 2011;4:73-6.  Back to cited text no. 36
    
37.
Yokogawa H, Kobayashi A, Sugiyama K. Clinical evaluation of a new donor graft inserter for Descemet's stripping automated endothelial keratoplasty. Ophthalmic Surg Lasers Imaging 2012;43:50.  Back to cited text no. 37
    
38.
Foster JB, Swan KR, Vasan RA, Greven MA, Walter KA. Small-incision Descemet stripping automated endothelial keratoplasty: A comparison of small-incision tissue injector and forceps techniques. Cornea 2012;31:42-47.  Back to cited text no. 38
    
39.
Tan D. Special DSAEK techniques for Asian eyes. DSEK What You Need to Know About Endothel. Keratoplasty. USA: Slack Incorporated; 2009.  Back to cited text no. 39
    
40.
Kobayashi-Busin DSAEK Glide 2013. Available from: http://www.asico.com/product/kobayashi-busin-dsaek-glide/. [Last accessed on 2013 Jul 15]  Back to cited text no. 40
    
41.
Gangwani V, Obi A, Hollick EJ. A prospective study comparing EndoGlide and Busin glide insertion techniques in descemet stripping endothelial keratoplasty. Am J Ophthalmol 2012;153:38-43.e1.  Back to cited text no. 41
    
42.
de Sanctis U, Aragno V, Brusasco L, Damiani F, Grignolo F. Pull-through insertion technique for descemet stripping automated endothelial keratoplasty (DSAEK): Graft survival and endothelial cell loss after 1 year in 100 eyes. J Transplant Technol Res S 2011;2:991-2161.  Back to cited text no. 42
    
43.
Terry MA, Saad HA, Shamie N, Chen ES, Phillips PM, Friend DJ, et al. Endothelial keratoplasty: The influence of insertion techniques and incision size on donor endothelial survival. Cornea 2009;28:24-31.  Back to cited text no. 43
    
44.
Macaluso C. J3760.1 DSEK/DSAEK Endothelial Graft Inserter; 2011. Available from: http://www.janach.it/newsdett.aspx?n=2836. [Last accessed on 2013 Jul 20]  Back to cited text no. 44
    
45.
Macaluso C. Closed-chamber pulling-injection system for donor graft insertion in endothelial keratoplasty. J Cataract Refract Surg 2008;34:353-6.  Back to cited text no. 45
    
46.
Macaluso C. Communication by Khan SN [electronic mail], Acquiring Image of Macaluso Injector, UTA; 20 Jun, 2013.  Back to cited text no. 46
    
47.
Tan D. Endoglide DSAEK Insertion and Procedure; 2009. Available from: http://www.angioedupro.com/Sharpoint/?seek=394. [Last accessed on 2013 Jul 23].  Back to cited text no. 47
    
48.
FDA-USA. FDA Approval of Neusidl Corneal Inserter by Fischer Surgical. Available from: http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?ID=36541. [Last accessed on 2013 Jul 20].  Back to cited text no. 48
    
49.
Surgical F. NCI Spatula by Fischer Surgical. Available from: http://www.fischersurgical.com/NCIAnimation Video 3-31-09.wmv. [Last accessed on 2013 Jul 20].  Back to cited text no. 49
    
50.
Hannush SB, Aldave A, Perry HD. Discover the Advantages of Endothelial Keratoplasty. Ophthalmol Manag 2012;16:48-53.  Back to cited text no. 50
    
51.
Terry MA, Straiko MD, Goshe JM, Shamie N, Shah A, Alqudah AA, et al. Endothelial keratoplasty: Prospective, randomized, masked clinical trial comparing an injector with forceps for tissue insertion. Am J Ophthalmol 2013;156:61-68.  Back to cited text no. 51
    
52.
Keramed, “Endoinjector.” Available from: http://www.keramed.com/endoinjector.html. [Last accessed on 2013 Jul 20].  Back to cited text no. 52
    
53.
FDA-USA. FDA Approval of Endoserter by Ocular Systems. Available from: http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpcd/classification.cfm?ID=3982. [Last accessed on 2013 Jul 20].  Back to cited text no. 53
    
54.
O. Systems, “Endoserter by Ocular Systems.” Available from: http://www.us.ocularsystemsinc.com/physicians/the-endoserter/. [Last accessed on 2013 Jul 20].  Back to cited text no. 54
    
55.
Ide T, Yoo SH, Kymionis GD, Perez VL, Goldman JM, O'Brien TP. Descemet-stripping automated endothelial keratoplasty (DSAEK): Effect of nontoxic gentian violet marking pen on DSAEK donor tissue viability by using vital dye assay. Cornea 2008;27:562-564.  Back to cited text no. 55
    
56.
Stoeger C, Holiman J, Davis-Boozer D, Terry MA. The endothelial safety of using a gentian violet dry-ink 'S' stamp for precut corneal tissue. Cornea 2012;31:801-803.  Back to cited text no. 56
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5], [Figure 6]
 
 
    Tables

  [Table 1]


This article has been cited by
1 TO STUDY THE EFFECTS OF INTRASTROMAL CORNEAL RING GEOMETRY AND SURGICAL CONDITIONS ON THE POSTSURGICAL OUTCOMES THROUGH FINITE ELEMENT ANALYSIS
SALMAN N. KHAN,PANOS S. SHIAKOLAS
Journal of Mechanics in Medicine and Biology. 2016; 16(07): 1650101
[Pubmed] | [DOI]
2 Finite Element Analysis of Descemetís Stripping Automated Endothelial Keratoplasty (DSAEK) Surgery Allograft to Predict Endothelial Cell Loss
Salman N. Khan,Panos S. Shiakolas
Current Eye Research. 2016; : 1
[Pubmed] | [DOI]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Introduction
Endothelial Allo...
Endothelial Glides
Busin Glide
Macaluso Inserter
Tan Endoglide
Endothelial Allo...
Neusidl Corneal ...
Endoinjector
Endoserter
Discussion
Acknowledgments
References